96
Calibration
GB
SUPREMATouch
WARNING!
Setting the bridge current deletes the calibration history of the sensor. The current calibration for
the selected measuring point will be reset.
To set the bridge current:
(1) Connect sensor simulation module.
(2) Enter the sensor type dependent current in the
Sensor Current reference value.
(3) Start the setting with
Start
.
(4) Wait until the
Status
field indicates completion of the process.
(5) End the process with
End
.
(6) Remove sensor simulation module.
(7) Connect desired sensor.
When the bridge current is set and the sensor has warmed-up properly, carry out the first calibra-
tion of the measuring point.
In order to avoid accidental damage and destruction of the sensors by an excessive
bridge current, the setting must be carried out using a corresponding sensor equivalent
network (see chapter 5.1 "Sensor Simulation Modules" or sensor simulator).
Summary of Contents for SUPREMATouch MCP 20
Page 1: ...MSAsafety com Operating Manual SUPREMATouch Fire and Gas Warning Unit Order No 10121863 05...
Page 8: ...MSAsafety com User Instruction Manual SUPREMATouch Fire and Gas Warning Unit...
Page 72: ...MSAsafety com Service and Maintenance Guide SUPREMATouch Fire and Gas Warning Unit...
Page 232: ...232 Dimensions GB SUPREMATouch 18 Dimensions 18 1 Rack...
Page 233: ...233 Dimensions GB SUPREMATouch 18 2 Rail mounted Modules MRO8 TS Module MRO16 TS Module...
Page 235: ...235 Dimensions GB SUPREMATouch MRO20 8 TS Module 1 3 2 69 90...
Page 236: ...236 Dimensions GB SUPREMATouch MRO20 16 TS Module 2 5 2 64 73 relay dependent 90...
Page 237: ...237 Dimensions GB SUPREMATouch MRC TS Module MGT40 TS Module...
Page 238: ...238 Dimensions GB SUPREMATouch MHD TS Module MAT TS Module...
Page 241: ...For local MSA contacts please visit us at MSAsafety com Because every life has a purpose...