26
Alarms
8 Alarms
Access the Remote Device dialogue as described on page 15. The Alarm Options
will be displayed.
1) Alarm Mode: 1
2) Dirty Alarm Level (ISO): 0
3) Dirty Alarm Level (NAS1638 / AS4059E-1): 0
4) Dirty Alarm Level (AS4059E-2):
1A/2B/3C/4D/5E/6F
The PML² includes two solid state relays for connections to an external circuit.
These can be arranged to function as follows:
1 Alarm Mode: 0
Selecting option 0 switches relays 1 and 2 always off.
2 Alarm Mode: 1
Selecting option 1 will configure relays 1 and 2 as described on page 28 - refer
to the examples given on the simple wiring diagrams. Clean alarm levels are set
in accordance with the instructions starting on page 21. Alarm mode 1 is used
in conjunction with the Continuous Sampling test type, enabling the Analyser to
operate continuously until the specified clean alarm level is achieved.
3 Alarm Mode: 2
Selecting option 2 arranges the relays to operate when the Clean and Dirty alarm
levels are reached/exceeded. Alarm mode 2 will normally be used in conjunction
with the Continuous test type (refer to page 20), but can be used with all other test
types as well.
Set the Clean and Dirty alarm levels by inputting the desired levels in to the Remote
Device dialogue on (page 15, also setting the appropriate result format to ISO, NAS
or AS4059, as described on page 21).