Moisture Sensor
25
7 Moisture Sensor
The PML² version fitted with the optional moisture sensor module allows both
measurement of % saturation of water in oil (Relative Humidity) and temperature.
These are displayed as RH % and °C on the main/test progress screen and on the
printed results.
Temperature measurement provides a reference temperature for the RH reading.
Due to the temperature gradient existing between the system tapping point
and the RH/temperature module, the temperature reading can be 5°C to 10°C
less than the actual system temperature, depending on operating conditions.
The PML² can be configured to do a test with or without the moisture sensor se-
lected. If the moisture sensor has been selected, the flush valve will open automat-
ically for a period of 3 minutes before the particle count test commences. This is
to allow the moisture sensor to stabilise and give an accurate reading.
Figure 1
Measure Water Content Enabled
To switch the moisture sensor on, select the Settings option in the Remote Device
dialogue, as described on page 15. The "Measure Water Content’’ box should be
ticked.