Remote Display - Optional Equipment
a guide, a 200ml flush volume is recommended when using the 1.5m long minimess
hose. Refer to page 36 for more information about flushing times.
After the flush volume has been discharged to waste the flush valve may be manu-
ally closed by again pressing pushbutton F. Alternatively, the flush valve will close
automatically when pushbutton Start (I) is pressed.
Note! For version PML²-W (moisture and temperature indication), every test is
automatically preceded by a 3 minute flush period, to allow the moisture sensor to
stabilise and give an accurate reading. The 3 minute period starts form the time the
Start pushbutton (I) is pressed. Therefore, a separate flush period is unnecessary,
unless a flush period greater than 3 minutes is required. For the PML²-W version
the moisture and temperature function can be switched off - refer to page 18 for
details.
•
Press Start Pushbutton
This will initiate the commencement of a test. Each test comprises a sampling
cycle that sizes and counts the particles in a 15ml volume of test fluid, followed by
an emptying cycle during which the 15ml of fluid is discharged to waste. At the
end of the emptying cycle the PML² will come to rest and the test results will be
displayed on the LCD screen.
•
Stop Pushbutton (O)
This optionally
aborts
the test at any point in the sampling/ emptying cycle. Please
note: The next test will start with an emptying cycle before the test commences.
10.2 Downloading the Results
After using Pushbutton Operation the test results can be downloaded.
The PML² has a memory capacity of approximately 600 tests. When this memory
is full, the earliest of the test records will be over-written by the new tests.
Connect the PML² to a PC running LPA-View and follow the instructions given to
Transfer Log on page 14.
It is important to set the Test Reference (refer to page 15) before transferring the
test results, as all results transferred to the databasewill be given the same Test
Reference that is active at the time in the Settings dialogue box.