36
Flush Valve Opening Times
13 Flush Valve Opening Times
The table below shows the flush valve opening times that are required to achieve a
200ml flush volume, for different conditions.
Model
Oil Inlet
Drain Pressure
Approximate Time
Type
Pressure
for 200ml
Flush Volume
Model A
2 bar
Atmospheric pressure, zero back pressure 2 minutes 30 seconds
Model B
10 bar
Back pressure not exceeding 0.5 bar
1 minute 20 seconds
Model B
10 bar
Back pressure not exceeding 1.0 bar
2 minutes
The above flush times assume that the oil inlet hose from the sampling point to the
PML² does not exceed 1.5m in length. Longer lengths of hose should be avoided, as
they can become a source of cross-contamination between tests. However, if a hose
length greater than 1.5m is used, the flushing time should be increased pro-rata.