PC Operation
13
•
Follow the installer Wizard, accepting the defaults.
•
When the Installer has completed, plug in the USB to Serial Adaptor.
•
A message should pop up indicating successful hardware instillation. Note any
COM port number indicated in the message.
•
Connect the PML² to the computer using the Serial lead and the USB to Serial
Converter.
•
If necessary, determin the COM port allocated by the computer for this device
using the proceedure above.
4.4 Operation
The following describes the function of the buttons on the Remote Device dialogue.
Figure 4
Remote Device dialogue
Start
Starts sampling and emptying cycle.
Stop
Stops test at any point in the sampling / emptying cycle. Next
test will start with an emptying cycle before the test commences.
Flush
The Flush Valve must be opened to allow fluid to pass through
the PML², so that any fluid from the previous test is flushed out
of the connecting pipe to eliminate the possibility of cross-con-
tamination between tests. The time for which the flush valve is
left open depends on the distance of the PML² from the system
being sampled and the conditions existing in the drain reservoir
- refer to page 36 for guidance. However, for version PML²-W