Appendix I
58
Specification
Typical test time
Result in <2.5 mins. (normal test)
Data storage
600 tests
Computer interface
RS 232 communication port
Hose connections
Minimess fittings
Microbore hose 1.5 metres long
Waste fluid hose
Dimensions
Height 120mm Length 277mm
Width 280mm Weight 5.9 kilos
IP Rating
IP66
Operation
Model A (Design Reference 30):
Minimum oil inlet pressure: 2 bar
Maximum oil inlet pressure: 400 bar
Drain Reservoir/System: atmospheric, zero back pressure.
Model B (Design Reference 31):
Minimum oil inlet pressure: 10 bar
Maximum oil inlet pressure: 400 bar
Drain Reservoir/System: Back pressure not exceeding 1
bar.
Analyser Upper Contamination Limmit
The analyser upper operating limit is set at 24/22/20. Tests that result in particle counts
exceeding any scale number in the three part ISO upper limit has the scale number
replaced by asterisk on the remote Display (optional equipment), example */*/*. Sim-
ilarly for NAS1638/AS4059E-1 and AS4059E-2 the Display will show ** and * re-
spectively.
Within LPA-View, exceeding the upper contamination level is denoted by the use of
code 25 in the case of ISO, and Class 15 for NAS1638/AS4059E-1 and AS4059E-2.