18
Normal Test
7. At the end of the Sampling cycle the test result will be displayed in the Remote
Device dialogue box.
If the Remote Device dialogue is open, the test results will be automatically trans-
ferred to the test database.
Following the sampling results the PML² discharges the sample fluid to waste. Test
status is shown as Emptying.
When emptying cycle is finished the status is shown as Ready, and another test can
then be started.
Figure 2
A Completed Test
5.2 With Moisture Sensor
For the PML²-W the procedure is similar to that detailed for the PML² without
moisture sensor (previous section), except when the Start button is pressed the
flush valve will open automatically for a period of 3 minutes, to allow the moisture
sensor to stabilise and give an accurated reading. ("Measure Water Content’’ box
in the Remote Device Setting dialogue has to be ticked to obtain a result).
Temperature is displayed in °C and moisture is expressed in % RH (Relative Hu-
midity).