Remote Display - Optional Equipment
31
W
Waiting (Continuous testing only)
C
Test Completed (Continuous testing only)
2 Pushbuttons
•
F – Black – Toggles the flush valve open and closed.
•
I – Green – Starts a test.
•
O – Red – Aborts a test.
3 Cleanliness Codes
ISO Code
– The test result is displayed as a three part code e.g. 22/19/16.
NAS1638/AS4059E-1 Class
– A single number is displayed, which is the high-
est recorded in all of the sizes, e.g. NAS 8.
AS4059E-2
– A single number is displayed, which is the highest record for all
of the sizes, e.g. AS 7.
10.1 Using the Pushbuttons
All new PML²’s are factory set to perform a Normal test, which consists of one
test sampling 15ml of fluid. Other test types can be selected using a PC running
LPA-View - refer to page 10.
When running a Normal test the user will first make the oil and power supply con-
nections as described on page 8.
Follow the actions below to perform a Normal test:
•
Press Flush Pushbutton (F)
This will open the flush valve to allow fluid to pass through the PML², so that any
fluid from the previous test is flushed out of the connecting pipe to eliminate the
possibility of cross-contamination between tests. The time for which the flush valve
is left open depends on the distance of the PML² from the system being sampled. As