22
Continuous Sampling
For continuous testing until the ISO Code is achieved, select ISO Format in the
Remote Device Settings dialogue, as described under the previous section on
page 15.
Testing will automatically continue until each of the three numbers in the Code
have been achieved (or cleaner).
Figure 2
Setting ISO Target Cleanliness Level
Clean Alarm Level (NAS1638 / AS4059E-1) - Input desired Clean Alarm Level,
as a single Class number in the range 2 to 12 inclusive.
For continuous testing until the (NAS1638 / AS4059E-1) Class is achieved,
select NAS Format or AS4059E Table 1, as described on page 15. Testing will
automatically continue until the Class number has been achieved at each of
the five micron size ranges covered by NAS 1638 & AS4059E Table 1. Note:
AS4059E-1 denotes Table 1 of the AS4059E standard.
Clean Alarm Level (AS4059E-2) - Input desired Clean Alarm Level in the for-
mat 1A/2B/3C/4D/5E/6F in the following range:
Size Code A: 000 to 12
Size Code B: 00 to 12
Size Code C: 00 to 12
Size Code D: 2 to 12
Size Code E: 4 to 12
Size Code F: 7 to 12