16
PC Operation
Dynamic
A comprehensive triple test
with results average: 30 ml sample volume
comprised of three 10ml sampling and emptying cycles. Allows the
effect of system fluctuations to be measured over a longer period of
time. The system
Triple / Bottle Sampling
A triple test with results average and quicker than the
Dynamic Test: 24ml sample volume comprised of three
individual 8 ml samples tested consecutively. For Bottle
Sampling refer to separate User Guide.
Continuous
For detailed instructions refer to Continuous Sampling, page 19.
Short
Single Test: 8 ml sample volume. This provides results in less time than
the Normal Test. It is not recommended for oil samples cleaner than ISO
17/15/12 (NAS 6), as the accuracy of the result might be compromised by
the ‘small’ sample volume
Test Number
Input desired number, e.g.’’123’’. (Test number will automatically
increment for each successive test).
Simulate Test Results
This is useful for checking device communications and
familiarisation with the software, especially when an oil
supply is not available. When a simulated test is performed,
the particle counter will generate test results without being
connected to a hydraulic system.
Format
Choose the preferred display format (ISO, Nas1638 / AS4059E-1, AS4059E-2).
This also sets which of the cleanliness targets is used for the continuous
test mode.
Serial Number
The serial number of the PML². This is recorded in each test re-
sult. (The serial number, together with the test timestamp, uniquely
identify the test record. These two parameters are used to avoid
duplication of test records.)
Software Revision
This identifies the program running on the PML².
Results will be displayed upon completion of three tests – including emptying cycle