Continuous Sampling
23
Example, 4A/4B/5C/6D/6E/7F.
For continuous testing until the AS4059E Table 2 size codes are achieved, select
AS4059E TABLE 2 FORMAT
, as described on page 15. Testing will automatically
continue until the Class number has been achieved at each of the six Size Codes.
Also, the PML² will handle deviations from the above format intelligently. The
size code can be out of order: 7F / 4A / 5C / 4B / 6E / 6D
If any sizes are missing, they will be assigned the “*” value. The effect of
this is a “don’t care” value when used as the cleanliness target. For example,
6B/6C/7D is translated as *A/6B/6C/7D/*E/*F. In this case, testing will con-
tinue until the B, C and D Classes are less than or equal to 6, 6, 7 respectively.
The A, E and F Classes are effectively ignored since they cannot ever be “
worse” than a “*” Class. Note: AS4059E-2 denotes Table 2 of the AS4059E
standard.
•
2Log all test results’’ box - A tick in this box will log every test that is carried
out on the continuous test mode, into the memory of the PML². If the box is
not ticked then the PML² will only store the results of the test when the Target
Cleanliness Level is achieved - this saves on memory space.
The PML² has a memory capacity of approximately 600 tests. When this mem-
ory is full, the earliest of the test records will be over-written by the newest tests.
•
Inserting a tick in the "Confirm Target Level’’ box instructs the PML² to repeat
the sampling cycle until the target cleanliness level has been achieved in two
consecutive samples, before the Complete status is displyed.
Leaving the "Confirm Target Level’’ box blank permits the target cleanliness
level to be achieved only one time before the Complete status is displayed.
•
Press OK button to save settings and return to Remote Device dialogue.
•
Press Flush button to open flush valve - flush indicator turns yellow to indicate
valve is open. Leave valve open until at least 200ml has passed to waste. Refer
to page 36 for more information about flushing times.
•
Press Stop button to close flush valve.
•
Press Start button, the PML² will now commence the sampling cycle.