External Circuit Relays
9 External Circuit Relays
Page 8, item b explains actions to be taken if it is intended to use the external circuit
relays that are built into the PML².
There are two solid state relays for external circuit use. The function of these re-
lays is illustrated in the following simple wiring diagram, using a battery and bulb
circuit for illustrative purposes.
Each relay is designed for a maximum current of 1 amp at 24 volt nominal AC
or DC (absolute maximum 60 volt peak). Operation above these limits will cause
irreparable damage to the relays.
If the User requires to switch voltages/currents in excess of the above maximum
limits, then separate higher rated interposing relays will need to be incorporated
into the final electrical scheme designed by the User.
In the multi-core cable that is supplied with the PML² there are three cores that
are used to connect with the relays. These cores are colour coded black, white and
yellow, see diagrams below.
The green core is not used and can be cut-off.
Figure 1
Example 1: Bulb illuminates when target
cleanliness level is achieved (Completed status) and
is off during sampling.