Section 4
4-13
System Printer(s)
Refer to the provided network printer manual for diagnostic testing, error logs, troubleshooting flow chart and
prescribed maintenance.
Receiver Module(s)
After verifying one of the receiver modules meets
specifications during the previous “Signal Strength Verification”
testing, the test transmitter should be mapped to each of the separate receiver modules installed in the system
hardware.
Each receiver module should be checked for proper operation, making sure that the signal strength level
measures >-80db and the ECG waveforms do not show signs of distortion or drop out with the transmitter placed
at location #1 from the previous test.
Transmitter(s)
Turn on the ECG simulator and connect it to one transmitter at a time from location #1 defined in the previous
tests. Start a monitoring session on the monitoring slot that matches the channel of the transmitter being tested
and verify the ECG quality is clean with out dropout and has a signal strength >-80db.
Repeat this test for all system transmitters.
S12/S19 Patient Monitors
Turn on the ECG simulator and connect it to one of the S12 or S19 Patient Monitors at a time Start a monitoring
session on the monitoring slot that matches the Bed ID and Unit ID of the S12 or S19 being tested and verify the
ECG quality is clean with out dropout.
Repeat this test for all system transmitters.