Mortara Instrument Inc.
Mortara Instrument, Inc.
Phone (414) 354-1600
7865 N. 86
th
Street
Fax (414) 354-4760
Milwaukee, WI. 53224
Surveyor Central Preventive Maintenance Report
System Serial #:
□ System/Network Verification
□
Verify network connectivity from node to node
□
Test export and archive functions
□
Check Time Synchronization
□ Audible Alarm Testing
□ UPS Testing / UPS Battery Verification - Check if N/A
□
UPS Testing
□
UPS Battery Verification
Replaced Date
/
/
or Not yet Required
(check)
□ System PC(s) Inspection/Cleaning
□ Antenna Network Maintenance
□
Coax Cable Check
□
Antenna Network Box Check
□
LNA/Down Converter Check
□
Antenna Network Power Supply Check
□ System Printer(s) - Check if N/A
□
Functional Check
□
Printer Cleaning
□ Signal Strength Verification
□
Antenna Network Verification
(attach covered area map w/locations)
Transmitter Location
Antenna A Power
Antenna B Power
1
2
3
4
5
6
□
System Receiver Module Verification
□
System Transmitter Verification
Performed By:
Date:
/
/
Summary of Contents for Surveyor Central 3.x
Page 4: ...iii...
Page 16: ...Surveyor Central 2 2 Note The Telemetry Nodes are computers that contain the PCI RF Boards...
Page 26: ......