Chapter 9: Regional Picking Processes
5031498 G
135
To undo all selections, click the Remove User Selections icon to display the Remove User
Selections dialog. Click Yes.
If you run a routine for white light and fluorescent imaging, two tabs are available.
The White Light tab displays the image taken with white light.
The other tab is labeled with the fluorescent filter pair and displays the image taken with
fluorescent light.
Switch between the two images to define the pickable colonies. This option is available only for
instruments with a fluorescence imaging module.
To select the colonies for regional picking:
1. From the list in the upper-right, select the barcode or identifier of the receptacle to view.
2. Refine the colony Selection criteria.
Compactness sets the level of irregularity for picking colonies. The value is a ratio of
the perimeter divided by the area of the colony, so that irregular shaped colonies are
closer to 0 and colonies that are more of a perfect circle are closer to 1. The system
default value is that a colony equal to or less than 0.65 and is not picked.
Axis Ratio measures how oval the colony is. The value is a ratio of the longest diameter
divided by the shortest diameter, so oval shaped colonies are closer to 0 and round
colonies are closer to 1. The system default value is that a colony equal to or less than
0.65 is not picked.
Min Diameter (mm) sets the minimum diameter of colonies for picking. A colony equal
to or smaller than the value in this field is not picked.
Note: The Min Diameter (mm) cannot be lower than the Diameter value set in the
Debris Discard section of the test image.
Max Diameter (mm) sets the maximum diameter of colonies for picking. A colony equal
to or greater than the value in this field is not picked.
Min Proximity (mm) sets the distance between colonies to be picked, so that when
picking one colony, a different adjacent colony is not picked. The system default value
is 0.45 mm.
Summary of Contents for QPix 450
Page 1: ...5031498 G August 2022 QPix 450 or QPix 460 Colony Picking System User Guide ...
Page 6: ...QPix 450 or QPix 460 Colony Picking System User Guide 6 5031498 G ...
Page 22: ...QPix 450 or QPix 460 Colony Picking System User Guide 22 5031498 G ...
Page 56: ...QPix 450 or QPix 460 Colony Picking System User Guide 56 5031498 G ...
Page 102: ...QPix 450 or QPix 460 Colony Picking System User Guide 102 5031498 G ...
Page 158: ...QPix 450 or QPix 460 Colony Picking System User Guide 158 5031498 G ...
Page 176: ...QPix 450 or QPix 460 Colony Picking System User Guide 176 5031498 G ...
Page 218: ...QPix 450 or QPix 460 Colony Picking System User Guide 218 5031498 G ...
Page 226: ...QPix 450 or QPix 460 Colony Picking System User Guide 226 5031498 G ...
Page 250: ...QPix 450 or QPix 460 Colony Picking System User Guide 250 5031498 G ...
Page 256: ...QPix 450 or QPix 460 Colony Picking System User Guide 256 5031498 G ...