QPix 450 or QPix 460 Colony Picking System User Guide
10
5031498 G
WARNING! HIGH VOLTAGE
Power off the instrument and disconnect the power
cord before you do maintenance procedures that require removal of a panel or cover
or disassembly of an interior instrument component.
Do not try to use the instrument until all covers are replaced.
To provide access for disconnecting power from the instrument, maintain a 66 cm (26 in.)
minimum clearance area on the right side of the instrument.
To protect against fire hazard, replace the fuses only with the same type and rating as the
original factory-installed fuses.
Ultraviolet (UV) Light Safety
The instrument is equipped with an automatically locking door that locks whenever you run a
process. The door prevents UV light from passing through during operation.
As a safety measure, if the door is open, an electromagnetic switch prevents the instrument
from running. Never tamper with this switch, as it serves two purposes:
It prevents the motors from running to reduce the potential of physical damage.
It disables the UV light to prevent the risk of damage from UV radiation.
Magnet Safety
Persons with external or implanted medical devices need to evaluate the risks related to these
devices before entering an area where the instrument is in use. Keep magnetic storage devices
or strips, such as hard drives and credit cards, away from the instrument.
Interactions with metallic objects can create pinch hazards.
External or Implanted Medical Device Safety
Motors and their related drives and cabling are sources of electromagnetic fields.
Persons with external or implanted medical devices must evaluate the risks related to these
devices before entering an area where the instrument is in use. Keep magnetic storage devices
or strips, such as hard drives and credit cards, away from the instrument.
WARNING!
Due to the presence of electromagnetic fields, if you wear an external or
implanted medical device, keep 305 mm (1 ft) away from the drive magnets.
Heat and Burn Safety
The instrument is fitted with a high-temperature halogen dryer. The casing can become hot
during the drying cycle.
WARNING!
The casing of the halogen dryer can become hot during the drying cycle.
Before touching this area, make sure that the dryer has had time to properly cool to a
safe temperature.
Summary of Contents for QPix 450
Page 1: ...5031498 G August 2022 QPix 450 or QPix 460 Colony Picking System User Guide ...
Page 6: ...QPix 450 or QPix 460 Colony Picking System User Guide 6 5031498 G ...
Page 22: ...QPix 450 or QPix 460 Colony Picking System User Guide 22 5031498 G ...
Page 56: ...QPix 450 or QPix 460 Colony Picking System User Guide 56 5031498 G ...
Page 102: ...QPix 450 or QPix 460 Colony Picking System User Guide 102 5031498 G ...
Page 158: ...QPix 450 or QPix 460 Colony Picking System User Guide 158 5031498 G ...
Page 176: ...QPix 450 or QPix 460 Colony Picking System User Guide 176 5031498 G ...
Page 218: ...QPix 450 or QPix 460 Colony Picking System User Guide 218 5031498 G ...
Page 226: ...QPix 450 or QPix 460 Colony Picking System User Guide 226 5031498 G ...
Page 250: ...QPix 450 or QPix 460 Colony Picking System User Guide 250 5031498 G ...
Page 256: ...QPix 450 or QPix 460 Colony Picking System User Guide 256 5031498 G ...