The cut-off frequency can be specified for the low pass filter.
Valid range: 0.1 2.5 Hz.
The available options depend on the measuring time.
[External supply]
When the load cells are connected to an external supply, it is possible to switch to ≤8 V, to
tune Sense voltage monitoring to the lower supply voltage.
Selection: below or equal 8 V (≤8 V), <above 8 V> (>8 V)
[Test mode]
With [absolute], the test value is determined when the test is called.
With [relative], the deviation from the initially stored test value is displayed; see
Chapter
.
[W&M]
Setting for legal-for-trade mode.
Note:
The transmitter of the series PR 5220 are not approved for calibration.
[Standstill time]
The parameters [Standstill time] and [Standstill range]are used to define the stability of
the scale (stable balance).
The input for the parameter [Standstill time] is expressed in seconds.
Valid range: 0.00 2 s
If the tome is set to "0" there is no check. The standstill time must not be less than the
measuring time.
[Standstill range]
As long as the weight fluctuations remain within this range, the device is determined to
be stable.
The input for the parameter [Standstill range] is expressed in "d."
Valid range: 0.01 10.00 dc.
[Tare timeout]
Timeout for a tare/zeroset command that cannot be executed (e.g. due to mechanical
instability of the scale, incorrect filter setting, resolution too high, standstill condition too
strict).
The input is done as seconds.
Valid range: 0.0 <2.5> 25 s.
At 0.0 s taring is only carried out when the scale is already stable.
[Zeroset range]
Define a ±range around the zero point determined by the dead load during calibration;
within this range
-
the displayed gross weight can be set to zero by pressing the zero-setting key (or by
a corresponding external command), and
-
automatic zero tracking is active.
Setting range: 0.00 10000.00 d
7 Commissioning
Transmitter Series PR 5220
EN-93
Minebea Intec