17.5.2
Principle
Principle
DC voltage, delta-sigma converter, ratiometric to the
load cell supply voltage
Conversion/measuring
time
5, 10, 20, 40, 80, 160, 320, 640, 960, 1200, 1600 ms
Digital filter
Can be activated, active 4th order (low pass)
Characteristic: Bessel, aperiodic, Butterworth, Tscheby-
scheff
Limit frequency
configurable
17.5.3
Accuracy and stability
Accuracy*
0.5 µV/e corr. to 3 mV for 6000 e
0.5 µV/e corr. to 5 mV for 10,000 e
Class III, according to OIML R76/EN45501
Min. measuring signal
(OIML)*
6,000 e: ≥0.25 mV/V
10,000 e: ≥0.42 mV/V
Linearity*
<0.002 %
Zero standstill error (TK
0
)*
<0.02 µV/K RTI; ≤0.004 %/10K at 1 mV/V
SPAN stability error (Tk
span
)*
< ±2 ppm/K
* with a measurement time of 160 ms.
17.5.4
Sensitivity
Sensitivity
0.5 µV/e @ 10,000 e (Cl. III) OIML R76
Max. resolution
7.5 million internal counts at 3 mV/V, not legal-for-trade
Measuring input (measu-
ring dead load)
0 max. 36 mV DC, symmetrical to zero
Dead load suppression
Max. measuring signal of 36 mV DC (dead load + range);
input/calibration via software
17.5.5
Connecting cables
Length of the connecting cable between junction box and instrument
Cable type
PR 6135, PR 6135A
Max. 500 m – length of the load cell cable
Length of the connecting cable between weighing platform and instrument
Cable type
LiYCY
max. 500 m
Transmitter Series PR 5220
17 Specifications
Minebea Intec
EN-228