As soon as the gross weight exceeds range 1, the next highest range with the next highest
scale interval becomes valid (1->2->5->10->20->50).
Note:
During calibration, the multi-interval function is always switched off.
Example:
Range mode: "Multi-interval"
Interval range 1: 0 1500 kg (when calibrating set scale interval: 1 kg)
Interval range 2: 1500 2900 kg (next highest scale interval: 2 kg)
7.13 Calibrating xBPI-scale
7.13.1
General information
The legal-for-trade application of PR 5220 with a xBPI-scale is not possible.
1.
Choose "Range mode" from the
-[Weighing point]- [Calib]- [Param] menu.
2.
Select "Multi-interval" and confirm.
3.
Setting interval range 1: Enter "1500 kg" for range limit 1.
4. Setting interval range 2: Enter "2900 kg" for range limit 2.
5.
Press the
softkey to exit and save calibration.
Transmitter Series PR 5220
7 Commissioning
Minebea Intec