4.4.7.3
Connecting a load cell with a 6-wire cable
Terminal
Port/color code
Description
SIGN. M+
+ Meas./gn
+ Signal (LC output)
SIGN. M-
- Meas./gy
- Signal (LC output)
SENSE S+
+ Sense/wh
+ sense
SENSE S-
- Sense/bk
- Sense
LC SUPPLY V+
+ Supply/rd
+ Supply (excitation)
LC SUPPLY V-
- Supply/bu
- Supply (excitation)
LC SUPPLY
Grounding
Screen (ground)
4.4.7.4
Connecting 2 8 load cells (650
Ω
) using a 6-wire connecting cable
Connections are made via junction box PR 6130/.. using connection cable PR 6135/.. or
PR 6136/... .
Load cell supply circuit
-
Load resistance of load cell circuit ≥75
, e.g. 8 load cells of 650
each
-
Supply voltage is fixed to 12 V DC and protected against short circuit.
For further technical data, see Chapter
.
Transmitter Series PR 5220
4 Device installation
Minebea Intec
EN-40
Summary of Contents for PR 5220 Series
Page 234: ......