16 Error messages
16.1 Error messages measuring circuit
The internal weighing electronics can generate error messages, which are shown on the
weight display.
VNC text
Error and possible cause
Remote dis-
play
Arith err
Internal arithmetic overflow:
-
Faulty calibration values.
Error 1
Overload
The measuring signal is higher than Max + (x d):
-
Wrong setting.
-
Too much weight on the scale.
Error 2
Ext. meas. device error
Measuring input open:
-
The measuring signal is higher than the permissible range of
36 mV.
-
Measuring cable is interrupted (cable break detection).
-
Other hardware defect.
Error 3
Value exceeds display
The weight value is not displayed:
-
Too many digits have been set.
Error 4
No sense voltage
No sense voltage:
-
Load cells not connected.
-
Sense line or supply line is interrupted.
-
Wrong polarity or sense voltage is low.
Error 6
Negative input
Negative measuring signal:
-
Wrong polarity of load cell signal.
-
Wrong polarity of supply voltage.
Error 7
No values from scale
Internal weighing point:
The measuring signal is higher than the permissible range of 36
mV. Cannot read weight values from ADC (analog-digital conver-
ter).
-
Error in weighing electronics board.
-
Defective load cell.
-
Cable break.
Error 9
16 Error messages
Transmitter Series PR 5220
EN-219
Minebea Intec