SPM address
Type of data
R/W
Function
B7
X56
X57
X58
X59
X60
BYTE
BOOL
BOOL
BOOL
BOOL
BOOL
R
R
R
R
R
R
Active status
Test mode active
Calibration active
Instrument is tared
Pendeo only: no standstill
Pendeo only: operation with a simulated LC
X64
BOOL
R/W
Standard application only: Read/write marker bit 1
X65
BOOL
R/W
Standard application only: Read/write marker bit 2
X66
BOOL
R/W
Standard application only: Read/write marker bit 3
X72
BOOL
R/W
Switch D11 to net weight.
X112
BOOL
W
Zero device.
X113
BOOL
W
Tare device
X114
BOOL
W
Reset the tare of the instrument
X115
BOOL
W
Start the test mode
X116
BOOL
W
Finish the test mode
X117
BOOL
W
Reset the power fail signal
X118
BOOL
W
Set the preset tare value as tare
X119
BOOL
W
Act. Store the gross weight in the preset tare memory (D31)
X121
BOOL
W
Reset error B19 = 0.
B16
USDINT
R
Exponent
Number of decimals
Example: 1.23 is displayed
Exponent: 2
B17
USDINT
R
Weight unit 1 = mg, 2 = g, 3 = kg, 4 = t, 5 = lb, 9 = oz
B18
USDINT
R
Scale interval (multi-interval/multiple range = d1 or e1)
B19
BYTE
R
Last weighing point error, see Chapter
.
B20
BYTE
R
Higher byte of product code (0x54)
B21
BYTE
R
Lower byte of product code (0x10)
B22
BYTE
R
Major part of version number (1.0)
B23
BYTE
BYTE
Minor part of version number (1.0)
D6
UDINT
R
Serial number (board number)
W14
INT
R
Counter will be increased for every measured value.
D8
DINT
R
Current gross weight
D9
DINT
R
Current net weight
13 SPM
Transmitter Series PR 5220
EN-213
Minebea Intec
Summary of Contents for PR 5220 Series