Technical data
Description
Data
Internal connection
Contact strip
External connection
2 x terminal, 6-pin
2 x terminal, 4-pin
Wire gauge max. 1.5 mm
2
Number of inputs/out-
puts
4 (CH1, CH2, CH3, CH4)
Input voltage
Logic 0: U
DC
= 0 to 5 V or open
Logic 1: U
DC
= 10 to 28 V
Passive, external power supply required
Input current
<7 mA @ 24 V
<3 mA @ 12 V
Protection against incorrect polarity
Input frequency
Max. 200 Hz (50% ratio)
Output
Change-over contact
Max. switching voltage:
U
DC
= 30 V/U
AC
= 24 V
Max. switching current:
1 A
Switching frequency
Max. 0.5 Hz
Potential isolation
Inputs:
Yes, via optocoupler
Outputs:
Free relay change-over contact
Cables
Screened
Connect cable screen to the device.
Cable length
Max. 50 m
Dimensions (L×W×H)
60×106×22 mm
Weight
70 g
Process Controller Maxxis 4 PR 5500
4 Device installation
EN-62
Minebea Intec