Fast transient disturbances (Burst)
EN 61000-4-4
2 kV
Peak voltages (surge) 1.2/50 µs
EN 61000-4-5
1/2 V
Conducted disturbances by high fre-
quency coupling (0.15…80 MHz)
EN 61000-4-6
10 V
Voltage dips
EN 61000-4-11
0/40/70%
20/200/500ms
Mains inputs
Mains failure link
EN 61000-4-11
20 ms
7.4.3
RF interference suppression
Interference emission
EN 55011, Group 1, Limit class A, for industrial
sectors
7.5 Mechanics
7.5.1
Design
Control cabinet housing made of stainless steel and aluminum, protection class IP65, rear
IP20.
7.5.2
Dimensions
Width
186/192/199 mm
Height
90/96/103 mm
Depth
approx. 203 mm (incl. screen clamping rail)
7.5.3
Weights
Net weight
2.2 kg
Shipping weight
3.1 kg
7.6 Use in legal metrology
7.6.1
Documentation relating to calibration on the CD included
The documents and manuals listed in the appendix (see Chapter
PR 5500 CD.
7.6.2
Further information
For information about the signiicance of the CAL switch to be sealed, see Chapter
and [Commissioning] - [Overwrite protection] - [CAL switch] in the PR 5500 operating
manual.
Ensure that the parameter settings (see [System setup] - [Weighing points] - [Internal
weighing point] - [Parameters] in the PR 5500 operating manual) selected comply with
the legal requirements as well as the provisions of EC type approval certiication.
The party installing the scale is responsible for selecting legally permissible settings.
These settings need to be checked.