Description
Data
Temperature coef. Zero
(Tk
o
)
<0.05 µV/K RTI,
<0.004%/10K @ 1 mV/V
Temperature coef. Vst
(Tk
span
)
<±4.0 ppm/K
Cable length
Max. 300 m with PR 6135
Cable type
6
‑
wire with screen for entire cable and screen for measure-
ment cables, e.g., PR 6135/..
Dimensions (LxWxH)
85×50×25 mm
Weight
Approx. 60 g
4.5.4.1
Analog connections
The connection for analog load cells or analog platforms (e.g. CAPP series) is located on
the housing rear panel. The supply voltage is protected against short circuit and overload.
Note:
Do not shorten
the load cell cable. Connect the prepared cable end and roll up the
remaining cable.
5
1
6
9
ST
OP
PROFIBUS DP
o
+
o
+ o
+
Terminal
Connection
Description
M+
+ Meas.
+ Signal/LC output
M-
- Meas.
- Signal/LC output
V+
+ Supply
+ Supply/excitation
S+
+ Sense
+ Sense
S-
- Sense
- Sense
V-
- Supply
- Supply/excitation
Process Controller Maxxis 4 PR 5500
4 Device installation
EN-52
Minebea Intec