Accutorr Plus™ Operating Instructions
3 - 37
Operation
Status and Error Codes
SYSTEM
984*
NIBP Hardware Failure
NIBP A/D failure detected.
985*
NIBP Overpressure Circuit
not Programmed
The overpressure circuit is not set to the
current patient size.
986*
NIBP Overpressure Circuit
not Tracking
The two pressure transducers are not
tracking each other.
987*
Stop - Hardware
Overpressure
Excessive cuff pressure detected by
hardware over-pressure sensor. A triple
beep tone is generated.
988*
TEMP Bad Calibration
Thermometer needs calibration.
990*
TEMP Illegal Mode
Thermometer switch is set wrong.
991*
TEMP Module Failed
Thermometer internal failure.
995*
SpO
2
Uncalibrated
SpO
2
fails calibration check.
996*
SpO
2
Failure
SpO
2
failed self-test.
Status and Error Code Table
TYPE
CODE
DESCRIPTION
REASON