Accutorr Plus™ Operating Instructions
3 - 11
Operation
Alarms
Alarm Limit Table
PARAMETER
RANGE
UNITS
FACTORY
DEFAULT
UNITS OF
INCREMENT
Systolic High
Adult
Pediatric
Neonate
Off, 60-260
Off, 60-160
Off, 50-125
mmHg
Off
5
Systolic Low
Adult
Pediatric
Neonate
Off, 55-150
Off, 55-130
Off, 45-115
mmHg
Off
5
Diastolic High
Adult
Pediatric
Neonate
Off, 40-200
Off, 40-150
Off, 35-100
mmHg
Off
5
Diastolic Low
Adult
Pediatric
Neonate
Off, 30-120
Off, 30-50
Off, 25-50
mmHg
Off
5
MAP High
Adult
Pediatric
Neonate
Off, 90-200
Off, 90-150
Off, 60-110
mmHg
Off
5
MAP Low
Adult
Pediatric
Neonate
Off, 40-100
Off, 40-70
Off, 30-70
mmHg
Off
5
Pulse Rate High
Adult
Pediatric
Neonate
Off, 100-245
Off, 100-245
Off, 100-245
bpm
Off
5
Pulse Rate Low
Adult
Pediatric
Neonate
Off, 35-120
Off, 35-150
Off, 75-200
bpm
Off
5
SpO
2
High
Adult
Pediatric
Neonate
Off, 61-99
Off, 61-99
Off, 61-99
%SpO
2
Off
1
SpO
2
Low
Adult
Pediatric
Neonate
60-95
60-95
60-95
%SpO
2
86
1