Status and Error Codes
Operation
3 - 36
Accutorr Plus™ Operating Instructions
3.16
Status and Error Codes
The Accutorr Plus uses the various displays on the front panel to display the operational
status. Status and error codes listed below can generally be resolved by the user however,
some error codes, which are marked with an asterisk (*), may require resolution by a
qualified technical service person. These codes with their descriptions are listed on the back
of the Quick Reference card.
NOTE:
Status codes 8810 through 8858 can be cleared from the
Room and Bed displays by pressing the Room/Bed key (24).
Status and Error Code Table
TYPE
CODE
DESCRIPTION
REASON
NIBP
8810
Retry - Unable to Measure
Motion artifact, cycle time-out, weak
pulsations or no pulsations. A triple beep
tone is generated.
8811
Retry - Pump Higher
Insufficient cuff pressure. A triple beep tone
is generated.
8812
Stop - Cuff Overpressure
Excessive cuff pressure detected by the
software. A triple beep tone is generated.
8813
Stop - Unable to Measure
4 successive measurement attempts failed.
A triple beep tone is generated.
TEMP
(PTM)
8830
Check Probe
Tissue contact may have been lost.
8831
Replace Probe
Defective probe or connection.
SpO
2
8850
No Sensor
No sensor connected.
8851
Sensor Off
Sensor not on patient. (Masimo SpO
2
only)
8852
Interference
Interference on signal. (Masimo SpO
2
only)
8853
Pulse Search
Unit cannot find signal. (Nellcor SpO
2
Module will report “Pulse Search” -8853-
when the sensor is not on the patient.)
8854
Weak Pulse
Weak pulse detected. (Masimo SpO
2
only)
8856
Check Sensor
Sensor problem. (Masimo SpO
2
only)
8857
PR<21
Pulse rate is less than 21 bpm. (Nellcor
SpO
2
only)
8857
PR<26
Pulse rate is less than 26 bpm. (Masimo
SpO
2
only)
8858
PR>249
Pulse rate is greater than 249 bpm. (Nellcor
SpO
2
only)
8858
PR>239
Pulse rate is greater than 239 bpm.
(Masimo SpO
2
only)