Electromagnetic Compatibility
Appendix
6 - 12
Accutorr Plus™ Operating Instructions
TABLE 6-3
GUIDANCE AND DECLARATION - ELECTROMAGNETIC IMMUNITY
The
Accutorr Plus
is intended for use in the electromagnetic environment specified below. The
customer or the user of the
Accutorr Plus
should assure that it is used in such an environment.
IMMUNITY
TEST
IEC 60601 TEST
LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC
ENVIRONMENT - GUIDANCE
Portable and mobile RF
communications equipment should be
used no closer to any part of the
Accutorr Plus
, including cables, than
the recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation
distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80
MHz
3 Vrms
d
= 1.2 x
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5
GHz
3 V/m
d
= 1.2 x
80 MHz to 800 MHz
d
= 2.3 x
800 MHz to 2.5 GHz
where
P
is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and
d
is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should be
less than the compliance level in each
frequency range.
b
Interference may occur in the
vicinity of equipment marked
with the following symbol:
NOTE:
At 80 MHz and 800 MHz, the higher frequency range
applies.
NOTE:
These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
the Accutorr Plus is used exceeds the applicable RF compliance level above, the Accutorr Plus should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the Accutorr Plus.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
P
P
P