Patient Setup and Room/Bed Assignment
Operation
3 - 4
Accutorr Plus™ Operating Instructions
NOTE:
The default patient size and initial cuff inflation pressure can
be customized. See section 3.15, “User Configuration” for
Room Number and Bed Letter
3.2.3
Room Number and Bed Letter
To monitor more than one patient, assign each patient to a particular room number and bed
letter. Use the ROOM/BED key (24) to set the room number from 0 to 99 and the bed letter
as a, b, c or d. On initial power up (no stored patient data), the room number and bed letter
default to 0,a.
FIGURE 3-2
Room Number and Bed Letter Keys and Indicators
1.
Press the
ROOM/BED
key (24). The ROOM LED flashes indicating that the room
number can now be changed.
2.
Press the
PATIENT INFO. Up or Down Arrow
key (27 & 28) to increment or
decrement the room number.
3.
Press the
ROOM/BED
key again. The BED LED flashes.
4.
Press the
PATIENT INFO.
Up or Down Arrow
key (27 & 28) to increment or
decrement the bed letter.
5.
Press the
ROOM/BED
key a third time to exit this mode, or do not press the key for 15
seconds.
Once measurements have been taken, and the unit is powered off and on, the room number
and bed letter will default to the lowest room and bed where data is currently stored.
HOLD TO CLEAR.
MEMORY FULL
PATIENT INFO.
ROOM
BED
ROOM
BED
DATA
SCAN
DELETE
INFO.
Press to increase or
decrease the Room
Number and Bed Letter
Press to change the
Room Number and
Bed Letter