Specifications
Appendix
6 - 2
Accutorr Plus™ Operating Instructions
6.2
Specifications
6.2.1
Systolic Pressure Readout
6.2.2
Diastolic Pressure Readout
*Tested per ANSI/AAMI SP10-1992, ANSI/AAMI SP10A-1996 methods
6.2.3
NIBP Measurement Cycle Time
Less than 30 seconds average at 80 BPM with 180mmHg pump up pressure, without retries,
motion artifact or arrhythmia with standard adult cuff on a healthy individual. Cycle time is
affected by arm size and wrapping technique.
6.2.4
Pulse Rate
6.2.5
Maximum Cuff Pressure
Two means of limiting cuff pressure are provided; a hardware over pressure monitor which
limits the pressure to 330mmHg for Adults, 220mmHg for Pediatrics, and 165mmHg for
Neonates. A software overpressure monitor which vents if the pressure exceeds 300mmHg
for Adults, 200mmHg for Pediatrics and 150mmHg for Neonates. If the hardware over
pressure circuit is tripped in normal operation then the unit must be turned off and back on to
reset the system.
Number of Digits:
3
Accuracy*:
Mean error <5 mmHg, Standard deviation
less than <8 mmHg
Range:
Adult Mode: 55 to 260 mmHg
Pediatric Mode: 55 to 160 mmHg
Neonatal Mode: 45 to 120 mmHg
Number of Digits:
3
Accuracy*:
Mean error less than ±5 mmHg,
Standard deviation less than ±8 mmHg
Range:
Adult Mode: 30 to 200 mmHg
Pediatric Mode: 30 to 150 mmHg
Neonatal Mode: 25 to 100 mmHg
Range:
35-245 BPM for Adult and Pediatric
70-245 BPM for Neonates
Display Resolution:
1 BPM
Accuracy:
±3 BPM or ±3%, whichever is greater