Specifications
Appendix
6 - 4
Accutorr Plus™ Operating Instructions
6.2.8
SpO
2
– Masimo
®
Performance Specifications
*Neonatal accuracy specifications are based upon testing the N-3000 and N-25 neonatal
sensors on healthy adult volunteers in induced hypoxia studies, in the range of 70 - 100%
Sp02. The specified accuracy also takes into account published literature which predicts
that there may be small difference in % Sp0
2
reported by the Oximetry when measurements
from adult and fetal blood 100% fetal hemoglobin are compared. Fetal hemoglobin is
present in concentrations varying from 10% to 90% in neonatal blood, and this percentage
declines over time. As the percentage of fetal hemoglobin in neonatal blood declines, the
theoretical effect on accuracy due to this source is reduced.
Range:
70-100% SpO
2
Display Resolution:
1% SpO
2
Display Update:
<4 secs
SpO
2
Accuracy Saturation during No Motion Conditions
1
:
Adults / Pediatrics:
70% to 100% ± 2
5
<70% unspecified
Neonates (LNOP/LNCS):
70% to 100% ± 3
SpO
2
Accuracy Saturation during Motion Conditions
6
:
Adults / Pediatrics
2
:
70% to 100% ± 3
<70% unspecified
Neonates
3
(LNOP):
70% to 100% ± 4
Neonates
3
(LNCS):
70% to 100% ± 3
<70% unspecified
SpO
2
Response Time:
10 secs (Display Averaging Time is not user-
selectable for Masimo. It will be set to the
default value of 8 seconds.)
NOTE:
This time was measured with post average time at 8
seconds.
Low Perfusion Performance
4
:
>0.02% Pulse Amplitude and %Transmission
>5%
Saturation (%SpO
2
) ±2 digits
Pulse ±3 digits
Wavelengths Emitted:
660nm and 905nm
Maximum Emitted Energy:
30mW at 50
μ
A pulsed