Front Panel
Controls and Indicators
2 - 6
Accutorr Plus™ Operating Instructions
22. DATA SCAN KEY
Press this key (1 beep tone) to view previous measurements for the selected patient on the
Accutorr Plus and to enable the Delete Info. key (Accutorr Plus NIBP only). The LED indicator
next to the key illuminates. On the Accutorr Plus NIBP, use the Patient Info. Up & Down Arrow
keys (27 & 28) to scroll through the stored measurements for the selected patient. On all
models of the Accutorr Plus, press and hold this key (2 beep tones, approx. 6 seconds) to
scan all of the rooms and beds for stored measurements. Press the
DATA SCAN
key again
to stop on a particular room/bed. Press the
DATA SCAN
key again to exit this view mode.
23. Data Scan Indicator
This LED indicator is illuminated when viewing prior data.
24. ROOM/BED NUMBER KEY
Press this key to change the displayed Room/Bed. After pressing this key use the Patient Info.
Up & Down Arrow keys (27 & 28) to change the Room/Bed. This key is also used when
selecting a User Configuration item.
25. Bed Letter Display
This display is used to show the current patient bed letter. It is also used to display status
codes for NIBP, SpO
2
and Temperature and to display User Configuration items.
26. Room Number Display
This display is used to show the current patient room number. It is also used to display status
codes for NIBP, SpO
2
and Temperature, indicates which alarm is being set (Hi or Lo), and
displays a User Configuration item.
27. PATIENT INFO. Down Arrow KEY
This key is used to decrement the alarm limits when they are shown on the LED displays and
to decrement the hours, minutes, month, day and year in the clock set mode. This key is also
used to change the Room/Bed, to scroll through previous data and to change initial inflation
pressure.
28. PATIENT INFO. Up Arrow KEY
This key is used to increment the alarm limits when they are shown on the LED displays and to
increment the hours, minutes, month, day and year in the clock set mode. This key is also
used to change the Room/Bed, to scroll through previous data and to change initial inflation
pressure.
29. SET ALARMS KEY
This key is used to select the NIBP and SpO
2
(Accutorr Plus advanced models only) alarms to
be changed. Repeated presses of this key sequences through the choices of Systolic Hi,
Systolic Lo, Diastolic Hi, Diastolic Lo, Map Hi, Map Lo, Pulse Rate Hi, Pulse Rate Lo, SpO
2
Hi
and SpO
2
Lo. After the last available parameter, the next press returns the unit to normal
operation. Once the desired parameter is flashing, use the Patient Info. Up & Down Arrow
keys (27 & 28) to increment or decrement the alarm values.