Accutorr Plus™ Operating Instructions
3 - 35
Operation
User Configuration
9c
Neonate Alarm Limits
Set the default alarm limit
values for a Neonate size
patient. See Section 3.5 for
details on setting alarm
limits.
OFF, except
SpO
2
low
which is 86
10a
Alarm Volume
Set the volume of an alarm
signal. The choices are: 1,
2, 3, 4 & 5. 5 is the loudest.
4
10b
SpO
2
Volume
Set the volume of the SpO
2
beep. The choices are: Off
(or ——), 1, 2, 3, 4 & 5. 5
is the loudest.
OFF
11
Continuous Print
Choices are ON or OFF. If
User Configuration #3 is set
to display graphics, the
choices are -l- or -O-.
OFF
12
Reset to Factory Defaults
To change all of the User
Configuration items back to
the Factory Defaults, while
in User Config. #12, press
and hold the
START NIBP
key for 3 seconds.
USER
CONFIGURATION
NUMBER
FUNCTION
DESCRIPTION
FACTORY
DEFAULT