Safety Designations
Introduction
Safety Designations
Safety designations per IEC 60601-1 Standard:
Type of protection against electric shock
Class 1 with internal electric power source.
Where the integrity of the external protective
earth (ground) in the installation or its
conductors is in doubt, the equipment shall be
operated from its internal electric power
source.
Degree of protection against electric shock
Monitor - Type B applied part.
NIBP - Type BF defibrillation protected
applied part.
SpO2 - Type BF protected applied part.
Supply Connection
100-120 VAC / 220-240 VAC
50-60 Hz; 0.85 / 0.5 A
11.1 VDC Lithium Ion Internal Battery
Mode of Operation
Continuous
Protection Against Hazard of Explosion
Not Protected (Ordinary)
Protection Against Ingress of Liquids
Meets the requirements specified by IEC
60601-1, clause 44.3 and IEC 60601-2-30:
Non Protected Equipment (IPX1) as specified
in EN 60529.
Degree of Electrical Connection Between
Equipment and Patient
Equipment designed for direct electrical and
non-electrical connection to the patient.
Degree of Mobility
Mobile and/or hand held