MacroVIEW
MM179-001 - Operator Manual
10
Environmental Conditions:
- Operating temperature: 15°C (59°F) – 30°C (86°F)
- Relative humidity: 10% - 80% without condensation
- Altitude: up to 2000m
1.4. Transport and storage conditions
- Temperature: –20°C up to +60°C (-4°F up to 140°F)
- Humidity: up to 80% (up to 50% at 40°C (104°F)) non-condensing
1.5. Warning information
REACH regulation (EC) N. 1907/2006 - Information on Substances of Very High Concern (SVHC).
The battery in the device contains SVHC substances:
- Lead, Number: CAS 7439-92-1, EC Number: 231-100-4.
The MacroVIEW system must only be connected to a 3-pole earthed main socket outlet (ground connection
protection must be provided).
Disconnect the main plug from the socket and turn off the system with the appropriate button (see chapter
3.8) before assembling the system, connecting the optional parts and cleaning.
Equipment must be connected to the main socket outlet, which is accessible and visible by the operator.
The use of naked flames that can cause accidental sparks in the areas adjacent to the unit is prohibited.
The instrument is supplied with replaceable fuses, but their replacement must be effected by trained and
authorized SERVICE personnel. If the system is still not properly working, contact the local authorized and
trained SERVICE representative or the manufacturer.
If the system is not working properly, contact your local representative or the manufacturer (always provide
the serial number of the unit, see label at the back of the unit).
All parts of the equipment and all the optional parts shall be exclusively supplied by the manufacturer.
In case of malfunction, shutdown the camera, the PC and the entire unit, then restart them again. If the
system is still not working properly, contact your local representative or the manufacturer.
In the event of malfunction and return:
1. Turn the camera off.
2. Turn off the PC.
3. Turn off the system with the ON/OFF button on the trolley (chapter 3.8).
4. Unplug the power supply cable from the main socket.
5. Disconnect the keyboard, barcode reader and anything else, if present.
6. Disconnect the battery (chapter 2.3.2).
7. Clean and disinfect all parts of the unit as described in chapter 6.
8. Disconnect the PC and place it in the appropriate packaging.
9. Put all parts of the equipment into the package supplied by the manufacturer (original package).
10. Send the unit to the manufacturer for repair (see chapter 7.5).
Using the equipment in ways other than those specified may void the warranty provided
by the manufacturer.
DO NOT open the camera box. If you do, this shall void the warranty.
It is very important to turn OFF the camera and the MacroVIEW PC before disconnecting
the equipment.
Milestone provides two USB keys for every unit, containing the MacroView 5 software in
digital format, the operating system and the relevant drivers. We strongly recommend
storing the USB keys in a safe place, as Milestone is not in a position to supply other USB
keys or copies thereof.
Summary of Contents for 372700
Page 2: ......
Page 13: ...MacroVIEW MM179 001 Operator Manual 13 Click Next Click Next Click Install...
Page 96: ...MacroVIEW MM179 001 Operator Manual 96 Press enter Press Ctrl C...
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......