GB
36
Application
8. With one of the controls
or
, set the desired pulse width (50 to 300
μ
S).
9. Brie
fl
y press the button
“Set”. The display
“ramp.” (ramp time) will illumi-
nate and the ramp time will blink on the display
.
10. Set the desired ramp time (1 to 10 sec.) with one of the controls
or
.
11. Brie
fl
y press the button
“Set”. The display
“con.” (continuous time) will
illuminate and the continuous time will blink on the display
.
12. Set the desired continuous time (1 to 20 sec.) with one of the controls
or
.
13. Brie
fl
y press the button
“Set”. The display
“rel.” (relax time) will illuminate
and the relax time will blink on the display
.
14. With one of the controls
or
, set the desired relax time (1 to 20 sec.).
15. Brie
fl
y press the button
“Set”.
This program has now been saved and can be used like a preset program (see
page 33 and 34).
In order to reset the self-de
fi
ned programs to the factory default, hold the but-
ton
“ ” with the device on for approximately 5 seconds until a long signal is
audible.
Factory settings for the programs:
TENS
Program
Duration
Frequency
Pulse width
U-01
30 min.
50 Hz
200
μ
s
U-02
30 min.
1.0 Hz
100
μ
s
EMS
Program
Duration Fre-
quency
Pulse
width
Ramp/
drop
Continu-
ous time
Relax
U-01
30 min.
80 Hz
150
μ
s
1 sec.
8 sec.
5 sec.
Delete de
fi
ned
programs