o p e r a t i n g
t h e
a n a l y z e r
s
Reinstall the sensor module. Place the sensor module on the alignment
pin and, using your thumb, push it straight back into position.
Reconnect the pump tubing to the valve module and sensor module.
Push the sampler down into the closed position. Close the access door.
Reinstall the reagent module. Press YES to
REPLACEMENT
COMPLETED?
. Reagents are automatically primed from the
reagent module. When priming is complete, the display indicates the
detection of each fluid with
PASS
, then automatically returns to the
REPLACE COMPONENTS
screen.
A
WARMING UP
message may appear for up to 30 minutes.
During this time, the EasyStat will not calibrate or analyze samples.