Sampler Removal
To remove the sampler from the sensor module, place the sensor
module on its side with the sampler side up. Using a coin or fingers,
remove the retaining screw from the sensor module. Remove the screw
mount (includes spring).
Rotate the sampler into the capillary mode position, then pull it out of
the sensor module.
Installation
Install the new sampler (includes sample probe and wiper) in the sen-
sor module. Insert the sampler alignment pin into the sensor module.
Rotate the sampler into the closed position.
Install the screw mount (includes spring) into the sampler. Replace the
retaining screw and tighten. DO NOT over tighten.