40
o p e r a t i n g
t h e
a n a l y z e r
4
HOME MENU
1 ANALYZE SAMPLE
2 ANALYZE QC
3 CALIBRATE
4 DAILY CLEANER
5 SECOND MENU
DAILY CLEANER
LIFT SAMPLER TO
USE CLEANER
Daily Cleaner
The fluid path
must
be cleaned with daily cleaning solution to
remove protein deposits. The analyzer will display
CLEANER
REQUIRED
if more than 10 samples have been analyzed and
more than 24 hours have elapsed since the last cleaning cycle.
Select
DAILY CLEANER
. The display shows
LIFT
SAMPLER TO USE CLEANER
. The sampler must be in
the syringe mode position. Lift the sampler until
PROBE IN
CLEANER?
appears. Insert the probe into the cleaning solution
and press YES to aspirate. Hold the container in place until
REMOVE
CLEANER RETURN SAMPLER
is displayed.
Push the sampler down into the closed position. The cleaning cycle is
followed by an automatic two-point calibration to prepare the analyzer
for sample analysis.
LIFT SAMPLER TO ANALYZE pH
PCO2 PO2 Hct Na+ K+ Ca++
is displayed when cali-
bration is successful.
Probe
Daily
Cleaner
s
Ca++ is replaced with Cl- when Cl- mode is selected.