Calculations
Temperature Corrections
The measured results are often corrected to the patient body tempera-
ture. The temperature corrected results more accurately reflect the
in
vivo
conditions. The EasyStat analyzer performs these calculations
using the following equations1:
∆
pH/
∆
T = 0.0065 (7.4 - pH) - 0.0147
∆
log PCO2 = 0.019
∆
log PO2 = 5.49 x 10
-
11 (PO23.88) + 0.071
where:
∆
T is expressed in °C
PCO2 and PO2 are expressed in mmHg
pH is expressed in pH units
∆
T
∆
T
9.72 x 10
-
9 (PO23.88) + 2.30