To replace EasyStat components, first select
REPLACE
COMPONENTS
from the
SECOND MENU
, then proceed with
component replacement as described in this section. If any replace-
ment procedure exceeds 20 minutes, the analyzer beeps and
NEED
MORE TIME?
is displayed. Press YES to return to the previous
screen. When all replacements are complete, press the HOME key
to return to the
HOME MENU
.
After any component replacement procedure, the
WARMING UP
message may appear for up to 30 minutes. During this time, the
EasyStat will not calibrate or analyze samples.
Failure to select the proper component from the
REPLACE
COMPONENTS
menu could result in multiple reliability issues,
including flow problems and/or slope, noise or drift errors.