37
o p e r a t i n g
t h e
a n a l y z e r
r
03
08
13
18
23
31
FEB
DEC
-2SD
7.414
+2SD
QC LEVEL 2
LOT# 00000000
MEAN 7.414
SD
0.008
DATE
AA
7.446
pH
7.382
s
The statistics reported above each chart require a minimum of 5
stored quality control results. The maximum number of stored results is
93 for each level.
Levey-Jennings Chart:
Levey-Jennings charts plot the measured quality control results for each
day. The calculated mean value is shown by the solid, vertical center
line. The +/- 2 SD values are shown by the dashed lines. The limits of
the charts are the mean +/- 4 SD. The Levey-Jennings charts plot data
for the last 31 days. (See example below.)