Prior to installation, a reagent module must be stored at room temper-
ature for a minimum of four hours. When the center of the indicator on
the back of the reagent module becomes dark, the module has been
exposed to excessive temperature and must not be used.
Place the new reagent module into the front of the analyzer. The
guide arrow must point to the right side of the reagent module.
Push the module straight back, then firmly to the right to lock it into
place against the valve module. Press YES to
REPLACEMENT
COMPLETED?
. The reagents are automatically primed from the
reagent module. When priming is complete, the display indicates the
detection of each fluid with
PASS
, then automatically returns to the
REPLACE COMPONENTS
screen.
The reagent module contains encoded information which is read by
the analyzer upon installation of the reagent module. This information
includes: reagent pH, PCO
2
, PO
2
, Hct, Na+, K+, Ca++, and Cl- val-
ues, and install by date of the reagent module.
Electrodes
The best indicator that an electrode (pH, PCO
2
, PO
2
, Na+, K+,
Ca++, Cl- or reference) requires replacement is when the EasyStat
displays
SLOPE
↑
/
↓
,
mV OUT OF RANGE
or
DRIFT
for
a
single
electrode. These messages generally indicate that the elec-
trode is not functioning correctly. For detailed information, refer to the
Troubleshooting chapter of this manual.
To remove or replace an electrode, select
ELECTRODES
from
the
REPLACE COMPONENTS
menu. Fluid is automatically
purged from the sample flow path. When the purge cycle is complete,
REPLACEMENT COMPLETED?
is displayed. Proceed with
electrode replacement.
Failure to select the proper component from the
REPLACE
COMPONENTS
menu could result in multiple reliability issues,
including flow problems and/or slope, noise or drift errors.
Summary of Contents for EasyStat
Page 1: ...EDIC EasyStat Operator s Manual...
Page 5: ...Operator s Manual EDIC EasyStat Ref 7415 003770 901R1 2015 07...
Page 10: ...4...
Page 14: ...8 Front View 2 Analyzer Layout...
Page 152: ...146...
Page 154: ...148...
Page 156: ......
Page 157: ......