![Medica EasyStat Operator'S Manual Download Page 119](http://html1.mh-extra.com/html/medica/easystat/easystat_operators-manual_1760909119.webp)
113
s p e c i f i c a t i o n s
r
Typical performance data: Capillary Mode
Within-run precision of duplicates (NCCLS EP-5) using whole blood and serum
PO2 PCO2
pH
Na+ K+
Ca++
Cl-
N
20 20 20 20 20 20 20
Mean
102 37.1 7.423 142.6 4.23 1.37 102.0
SD
0.9 0.57 0.008 0.67 0.020 0.010 1.00
CV
0.9 1.5
0.5 0.5 0.7 1.0
All tables: pH is in pH units, PCO2 and PO2 are in mmHg, CV are in %.