t r o u b l e s h o o t i n g
g
s
•
Clean and dry all sensor module surfaces and electrode pin
receptacles.
•
Replace reference electrode if it has been installed longer than six
months or the red ball is not floating.
•
Wipe dry all electrodes before reinstallation.
•
Reinstall the electrodes and confirm that they are correctly
aligned. Each electrode is keyed to assist with correct placement.
Release the compression plate.
•
Close the access door.
•
Calibrate.
A
WARMING UP
message may appear for up to 30 minutes.
During this time, the EasyStat will not calibrate or analyze samples.