56
o p e r a t i n g
t h e
a n a l y z e r
Reconnect the pump tubing to the valve module and sensor module.
Push the sampler down into the closed position.
Press YES to
REPLACEMENT COMPLETED?
. Reagents
are automatically primed from the reagent module. When prim-
ing is complete, the display indicates the detection of each
fluid with
PASS
, then automatically returns to the
REPLACE
COMPONENTS
screen.
A
WARMING UP
message may appear for up to 30 minutes.
During this time, the EasyStat will not calibrate or analyze samples.
If the sampler was replaced, the sampler switches may be checked for
proper function. Select the
SAMPLER SWITCHES
test outlined
under Diagnostics in Operating the Analyzer.