24
o p e r a t i n g
t h e
a n a l y z e r
yyes
y
1
no
SAMPLE # XXX
ANALYZING SAMPLE
SAMPLE # XXX
PROBE IN SAMPLE?
SAMPLE # XXX
REMOVE SAMPLE
RETURN SAMPLER
LIFT SAMPLER TO
LIFT SAMPLER TO
ANALYZE
2
2
Hct Na
+
K
+
Ca
++
yyes
y
HOME MENU
1 ANALYZE SAMPLE
2 ANALYZE QC
3 CALIBRATE
4 DAILY CLEANER
5 SECOND MENU
no
SAMPLE # XXX
ASPIRATING...
CALCULATED VALUES
SCREENS
RETURN SAMPLER
ONE MOMENT...
ANALYSIS RESULTS #001
pH
7.403
PCO
2
32.5
mmHg
PO
2
118
mmHg
Hct
45
%
Na
+
140.1
mmol/L
K
+
4.02
mmol/L
Ca
++
1.33
mmol/L
1
s
Ca++ is replaced with Cl- when Cl- mode is selected.
Analyze Sample
The ACCESS DOOR must be closed during sample analysis.
Select
ANALYZE SAMPLE
. The display shows
LIFT
SAMPLER TO ANALYZE pH PCO2 PO2 Hct Na+
K+ Ca++
. When an electrode does not calibrate successfully, that
parameter will not be displayed on the screen. No sample results
(including dependent calculated results) are given for that electrode.
Refer to ELECTRODES in Troubleshooting.