142
142
t r o u b l e s h o o t i n g
Electronic failure messages
•
Power the analyzer off and then on. If the error contin-
ues, retain the printouts with the errors and contact your
EasyStat dealer.
Room temperature or barometric pressure not within
the operating specifications (15-30°C, 500-800 mmHg)
•
Check the room temperature or barometric pressure
by proceeding to
DIAGNOSTICS
and selecting
SENSOR STATUS
.
•
Move the analyzer to a location where the temperature
and barometric pressure fall within the operating limits.
•
If the error continues, contact your EasyStat dealer.
Incorrect values for EasyStat.
•
Only occurs when using barcode.
•
Scan correct barcode. If problem continues, contact your
EasyStat dealer.
HARDWARE
A/D ERROR
DS NOT UPDATED
EEPROM ERROR
MISMATCH ERROR
CLOCK ERROR
HARDWARE
TEMP/BP
↑
/
↓
QUALITY CONTROL
QC ERROR 1